Fluke Calibration 5627A-9-B

$475.00

In stock

Compare
SKU: 3003720 Category: Tags: ,
  • Vibration and shock resistant
  • 3/4-inch bend radius for increased durability
  • Resistance: Nominal 100 Ω
  • Temperature Coefficient: 0.00385Ω/Ω/ °C nominal
  • Temperature Range: -200 °C to 420 °C (5627-6 and 5627-9 to 300 °C, transition and cable temperature: 0 °C to 150 °C)
  • Drift Rate: ±0.13 °C at 0 °C after 1000 hours at 400 °C
  • Sheath Material: 316 Stainless Steel
  • Leads: Teflon™-insulated, nickel-plated stranded copper, 22 AWG
  • Termination: Bare Wire
  • Time Constant: Four seconds maximum for 63.2 %Response to step change in water moving at 3 fps.
  • Bending Radius: Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip.
  • Immersion: At least 100 mm (4 in) recommended
  • Accuracy: (includes calibration uncertainty and short-term stability)
  • ±0.050 °C at -196 °C ±0.050 °C at 0 °C ±0.051 °C at 200 °C
  • ±0.055 °C at 420 °C
  • Size: 9 in L x 3/16 in Dia.

>

Each Unit Includes

  • Accredited Calibration

Additional information

Brand

Calibration Fluke

Main Menu

Fluke Calibration 5627A-9-B